Introduction
TA2100 a wide-range, multi-value and high-precision calibrator, can accurately output ultra-precision AC voltage, current and power, with two specifications of 0.02% and 0.05%.
It is suitable for calibrating AC power analyzers, AC power meters, AC voltmeters, and AC ammeters.
Features
Electrical Parameters Output: Voltage is up to 1100 V, current is up to 110 A at a frequency of 45 Hz ~ 70 Hz (400 Hz optional).
AC power output: calibrate the active power, reactive power, apparent power and power factor of the meter under test.
Value Adjustment Method: Fixed-point, rotary encoder, step and potentiometer (option).
Output Switch: The output can be turned on/off arbitrarily by one-touch operation.
Statistical analysis function: Automatically record the maximum and minimum values of electricity, calculate the average value and standard deviation, and analyze the stability.
High reliability of the whole machine: with overload protection, voltage short circuit protection, current open circuit protection, overheating protection, etc.
Excellent Human-Computer Interaction: Large LCD touch screen display, touch and button operation modes, easy for users to measure manually.
Introduction
TA2100 a wide-range, multi-value and high-precision calibrator, can accurately output ultra-precision AC voltage, current and power, with two specifications of 0.02% and 0.05%.
It is suitable for calibrating AC power analyzers, AC power meters, AC voltmeters, and AC ammeters.
Features
Electrical Parameters Output: Voltage is up to 1100 V, current is up to 110 A at a frequency of 45 Hz ~ 70 Hz (400 Hz optional).
AC power output: calibrate the active power, reactive power, apparent power and power factor of the meter under test.
Value Adjustment Method: Fixed-point, rotary encoder, step and potentiometer (option).
Output Switch: The output can be turned on/off arbitrarily by one-touch operation.
Statistical analysis function: Automatically record the maximum and minimum values of electricity, calculate the average value and standard deviation, and analyze the stability.
High reliability of the whole machine: with overload protection, voltage short circuit protection, current open circuit protection, overheating protection, etc.
Excellent Human-Computer Interaction: Large LCD touch screen display, touch and button operation modes, easy for users to measure manually.
Specification
AC Voltage Output ACV
Range | Resolution | Stability (%/min) | Accuracy | Max. Load Current (mA) | Distortion (%) | ||
0.05% | 0.02% | 0.05% | 0.02% | ||||
3 V | 10 μV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 600 | < 0.5 |
10 V | 100 μV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 600 | < 0.5 |
30 V | 100 μV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 600 | < 0.5 |
100 V | 1 mV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 500 | < 0.5 |
300 V | 1 mV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 150 | < 0.5 |
600 V | 1 mV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 80 | < 0.5 |
1000 V | 10 mV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 50 | < 0.5 |
AC Current Output ACI
Range | Resolution | Stability (%/min) | Accuracy | Max. Load Voltage (mA) | Distortion (%) | ||
0.05% | 0.02% | 0.05% | 0.02% | ||||
3 mA | 10 nA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 | < 0.5 |
10 mA | 100 nA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 | < 0.5 |
30 mA | 100 nA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 | < 0.5 |
100 mA | 1 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 | < 0.5 |
300 mA | 1 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 | < 0.5 |
1 A | 10 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 | < 0.5 |
3 A | 10 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 30 | < 0.5 |
10 A | 100 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 6 | < 0.5 |
30 A | 100 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 3 | < 0.5 |
100 A | 1 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 1.5 | < 0.5 |
Output Range: 0.3 mA ~ 110 A; Resolution: 0.001 %*RG;
Open circuit protection, overload protection
AC Small Voltage Output
Range | Resolution | Stability (%/min) | Accuracy | Distortion (%) | ||
0.05% | 0.02% | 0.05% | 0.02% | |||
5 mV | 10 nV | 0.02 | 0.02 | 300 + 2 | 120 + 1 | < 0.5 |
20 mV | 100 nV | 0.01 | 0.01 | 300 + 4 | 120 + 2 | < 0.5 |
50 mV | 100 nV | 0.01 | 0.005 | 300 + 10 | 120 + 5 | < 0.5 |
200 mV | 1 μV | 0.01 | 0.005 | 300 + 40 | 120 + 15 | < 0.5 |
500 mV | 1 μV | 0.01 | 0.005 | 300 + 100 | 120 + 40 | < 0.5 |
2 V | 10 μV | 0.01 | 0.005 | 300 + 400 | 120 + 150 | < 0.5 |
5 V | 10 μV | 0.01 | 0.005 | 300 + 1000 | 120 + 400 | < 0.5 |
Output Range: 0.5 mV ~ 6 V; Resolution: 0.001 %*RG;
Short circuit protection, overload protection; This output is suitable for the small signal input of the sensor of the power analyzer.
AC Power Output
Stability (%/min) | Accuracy (k = 2) %*FS | |||
0.05% | 0.02% | 0.05% | 0.02% | |
0.01 | 0.005 | 0.05 | 0.02 | |
Reactive Power | 0.01 | 0.01 | 0.1 | 0.05 |
Apparent Power | 0.02 | 0.01 | 0.05 | 0.02 |
Power Factor | 0.02 | 0.01 | 0.1 | 0.02 |
Output Range of AC Power: Combination of AC voltage range and AC current range.
Range of Power Factor: -1.0000…0.0000…+1.0000
Active Power (|cosφ| ≥0.5), Reactive Power (|sinφ| ≥0.5)
Specification
AC Voltage Output ACV
Range | Resolution | Stability (%/min) | Accuracy | Max. Load Current (mA) | Distortion (%) | ||
0.05% | 0.02% | 0.05% | 0.02% | ||||
3 V | 10 μV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 600 | < 0.5 |
10 V | 100 μV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 600 | < 0.5 |
30 V | 100 μV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 600 | < 0.5 |
100 V | 1 mV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 500 | < 0.5 |
300 V | 1 mV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 150 | < 0.5 |
600 V | 1 mV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 80 | < 0.5 |
1000 V | 10 mV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 50 | < 0.5 |
AC Current Output ACI
Range | Resolution | Stability (%/min) | Accuracy | Max. Load Voltage (mA) | Distortion (%) | ||
0.05% | 0.02% | 0.05% | 0.02% | ||||
3 mA | 10 nA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 | < 0.5 |
10 mA | 100 nA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 | < 0.5 |
30 mA | 100 nA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 | < 0.5 |
100 mA | 1 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 | < 0.5 |
300 mA | 1 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 | < 0.5 |
1 A | 10 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 | < 0.5 |
3 A | 10 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 30 | < 0.5 |
10 A | 100 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 6 | < 0.5 |
30 A | 100 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 3 | < 0.5 |
100 A | 1 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 1.5 | < 0.5 |
Output Range: 0.3 mA ~ 110 A; Resolution: 0.001 %*RG;
Open circuit protection, overload protection
AC Small Voltage Output
Range | Resolution | Stability (%/min) | Accuracy | Distortion (%) | ||
0.05% | 0.02% | 0.05% | 0.02% | |||
5 mV | 10 nV | 0.02 | 0.02 | 300 + 2 | 120 + 1 | < 0.5 |
20 mV | 100 nV | 0.01 | 0.01 | 300 + 4 | 120 + 2 | < 0.5 |
50 mV | 100 nV | 0.01 | 0.005 | 300 + 10 | 120 + 5 | < 0.5 |
200 mV | 1 μV | 0.01 | 0.005 | 300 + 40 | 120 + 15 | < 0.5 |
500 mV | 1 μV | 0.01 | 0.005 | 300 + 100 | 120 + 40 | < 0.5 |
2 V | 10 μV | 0.01 | 0.005 | 300 + 400 | 120 + 150 | < 0.5 |
5 V | 10 μV | 0.01 | 0.005 | 300 + 1000 | 120 + 400 | < 0.5 |
Output Range: 0.5 mV ~ 6 V; Resolution: 0.001 %*RG;
Short circuit protection, overload protection; This output is suitable for the small signal input of the sensor of the power analyzer.
AC Power Output
Stability (%/min) | Accuracy (k = 2) %*FS | |||
0.05% | 0.02% | 0.05% | 0.02% | |
0.01 | 0.005 | 0.05 | 0.02 | |
Reactive Power | 0.01 | 0.01 | 0.1 | 0.05 |
Apparent Power | 0.02 | 0.01 | 0.05 | 0.02 |
Power Factor | 0.02 | 0.01 | 0.1 | 0.02 |
Output Range of AC Power: Combination of AC voltage range and AC current range.
Range of Power Factor: -1.0000…0.0000…+1.0000
Active Power (|cosφ| ≥0.5), Reactive Power (|sinφ| ≥0.5)
Frequency and Phase
Range of Frequency: 45.000 0 Hz ~ 70.000 0 Hz
Resolution of Frequency: 0.000 1 Hz
Accuracy of Frequency: 0.005 Hz
Range of Phase: 0.000 ° ~ 359.999 °
Resolution of Phase: 0.001 °
Accuracy of Phase: 0.05 ° (0.05%), 0.02 ° (0.02%)
Frequency and Phase
Range of Frequency: 45.000 0 Hz ~ 70.000 0 Hz
Resolution of Frequency: 0.000 1 Hz
Accuracy of Frequency: 0.005 Hz
Range of Phase: 0.000 ° ~ 359.999 °
Resolution of Phase: 0.001 °
Accuracy of Phase: 0.05 ° (0.05%), 0.02 ° (0.02%)