Introduction
TD1310, a high-precision AC/DC standard instrument, can measure AC/DC voltage, current and small signal voltage. The best accuracy is 0.005% (DC) and 0.01% (AC).
The instrument is also a high-level AC/DC standard power / energy meter, and the best measurement accuracy is 0.01% (DC) and 0.02% (AC).
l The application of this instrument is very wide, and it can directly verify or calibrate equipment such as AC/ DC voltage sources, current sources, power sources, and multi-function calibrators.
It can also be used as a standard meter with test power supply to calibrate voltmeter, ammeter, resistance meter, power meter, frequency meter, phase meter, transmitter, sensor and other measuring instruments.
In the field of electric energy measurement, the instrument can be used as a high-level electric energy standard meter to check AC/DC electric energy meter verification device, standard electric energy meter, charger tester and so on.
The measurement mode of DC + AC is especially suitable for the verification and calibration of DC energy meter verification device with ripple detection function.
Features
l Wide range of direct current measurement: a pair of terminals can be used to measure the DC current of 10 μA ~ 120A, which solves the top problem of automatic range switching of ultra-wide range DC current.
Current expansion measurement: it is equipped with a special measurement interface for the sensor, which can be connected with a through current sensor to expand the measurement range to 500 A, 1 kA, etc.
The interface can supply power for the sensor and measure the secondary signal at the same time. The primary current can be displayed directly, by setting the ratio of the sensor on the instrument.
Small signal voltage measurement: it can withstand the overload signal of 100V peak voltage. It is suitable for testing indirect access watt hour meter verification device. It can also be connected with shunt or I / V conversion standard to measure large current.
Statistics and Analysis of Data: Calculate the electrical parameters such as the maximum, minimum, average, difference, standard deviation, etc.
Multiple measurement modes: DC, AC, DC + AC three modes, fundamental frequency: 40Hz ~ 1kHz, and measure 2 ~ 63th harmonic and DC ripple.
High reliability of the instrument: complete electrical isolation between voltage and current measuring circuits.
When measuring real load power, the current measurement circuit can be placed on the high or low side of the voltage circuit.
Electric Energy Metering: Standard I/O pulse interfaces, used for building a standard for high-class AC/DC electric energy metering.
Excellent Human-computer Interaction: LCD touch screen, more intuitive of value display and easier operation. it is equipped with RS232, USB, RJ45 interfaces, which is convenient to set up a full-automatic measurement system.
Introduction
TD1310, a high-precision AC/DC standard instrument, can measure AC/DC voltage, current and small signal voltage. The best accuracy is 0.005% (DC) and 0.01% (AC).
The instrument is also a high-level AC/DC standard power / energy meter, and the best measurement accuracy is 0.01% (DC) and 0.02% (AC).
l The application of this instrument is very wide, and it can directly verify or calibrate equipment such as AC/ DC voltage sources, current sources, power sources, and multi-function calibrators.
It can also be used as a standard meter with test power supply to calibrate voltmeter, ammeter, resistance meter, power meter, frequency meter, phase meter, transmitter, sensor and other measuring instruments.
In the field of electric energy measurement, the instrument can be used as a high-level electric energy standard meter to check AC/DC electric energy meter verification device, standard electric energy meter, charger tester and so on.
The measurement mode of DC + AC is especially suitable for the verification and calibration of DC energy meter verification device with ripple detection function.
Features
l Wide range of direct current measurement: a pair of terminals can be used to measure the DC current of 10 μA ~ 120A, which solves the top problem of automatic range switching of ultra-wide range DC current.
Current expansion measurement: it is equipped with a special measurement interface for the sensor, which can be connected with a through current sensor to expand the measurement range to 500 A, 1 kA, etc.
The interface can supply power for the sensor and measure the secondary signal at the same time. The primary current can be displayed directly, by setting the ratio of the sensor on the instrument.
Small signal voltage measurement: it can withstand the overload signal of 100V peak voltage. It is suitable for testing indirect access watt hour meter verification device. It can also be connected with shunt or I / V conversion standard to measure large current.
Statistics and Analysis of Data: Calculate the electrical parameters such as the maximum, minimum, average, difference, standard deviation, etc.
Multiple measurement modes: DC, AC, DC + AC three modes, fundamental frequency: 40Hz ~ 1kHz, and measure 2 ~ 63th harmonic and DC ripple.
High reliability of the instrument: complete electrical isolation between voltage and current measuring circuits.
When measuring real load power, the current measurement circuit can be placed on the high or low side of the voltage circuit.
Electric Energy Metering: Standard I/O pulse interfaces, used for building a standard for high-class AC/DC electric energy metering.
Excellent Human-computer Interaction: LCD touch screen, more intuitive of value display and easier operation. it is equipped with RS232, USB, RJ45 interfaces, which is convenient to set up a full-automatic measurement system.
Specification
AC/DC Voltage Measurement
Range | Resolution | Accuracy (k = 2) ppm*RD +ppm*RG or V | Temperature Coefficient (/℃) | ||
DC | 40 Hz≤F≤200 Hz | 200Hz≤F≤1 kHz | |||
100 mV | 0.1 μV | 50 + 5 μ | 120 + 80 | 200 + 100 | 3 + 0.5 μ |
300 mV | 0.1 μV | 30 + 6 μ | 120 + 80 | 200 + 100 | 3 + 0.6 μ |
1 V | 1 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 1 |
3 V | 1 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 1 |
10 V | 10 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 0.5 |
30 V | 10 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 1 |
100 V | 100 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 1 |
300 V | 100 μV | 30 + 20 | 120 + 80 | 200 + 100 | 3.5 + 1.5 |
1000 V | 1 mV | 30 + 20 | 120 + 80 | 200 + 100 | 3.5 + 1.5 |
l Measuring Range of Voltage: ± (10 mV ~ 1150 V)
Range Switching: Automatic or Manual
Input Resistance: 1 MΩ
Range | Resolution | Accuracy (k = 2) ppm*RD +ppm*RG | Temperature Coefficient (/℃) | ||
DC | 40 Hz≤F≤200 Hz | 200Hz≤F≤1 kHz | |||
100 μA | 0.1 nA | 50 + 5 nA | 200 + 100 | 250 + 150 | 3 + 0.5 n |
300 μA | 0.1 nA | 50 + 15 nA | 200 + 100 | 250 + 150 | 3 + 1.5 n |
1 mA | 1 nA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
3 mA | 1 nA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
10 mA | 10 nA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
30 mA | 10 nA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
100 mA | 100 nA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
300 mA | 100 nA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
1 A | 1 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
3 A | 1 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
10 A | 10 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
30 A | 10 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
100 A | 100 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
200 A① | 100 μA | 30 + 20 | 200 + 100 | 250 + 150 | 3.5 + 1.5 |
500 A① | 100 μA | 30 + 20 | 200 + 100 | 250 + 150 | 3.5 + 1.5 |
Note: ①: Range of 200 A and 500 A are optional, which needs to be equipped with current sensor such as TH5450-500 A to realize range expansion
Measuring Range: ± (10 μA ~ 120 A)
Range Switching: Automatic or Manual
Direct input
DC Small signal voltage measurement DCIU (Indirect access)
Range | Resolution | Accuracy (k = 2) ppm*RD +ppm*RG | Temperature Coefficient (/℃) | ||
DC | 40 Hz≤F≤200 Hz | 200Hz≤F≤1 kHz | |||
1 mV | 1 nV | 40 + 0.2 μV | —— | —— | 4 + 0.02 μ |
10 mV | 10 nV | 40 + 0.5 μV | 120 + 80 | 200 + 100 | 3 + 0.05 μ |
100 mV | 100 nV | 40 + 10 | 80 + 40 | 100 + 50 | 3 + 0.5 |
1 V | 1 μV | 40 + 10 | 80 + 40 | 100 + 50 | 2 + 0.5 |
10 V | 10 μV | 40 + 10 | 80 + 40 | 100 + 50 | 2 + 0.5 |
Measuring Range: ± (0.1mV ~ 11 V)
Input Resistance: >1 GΩ
Input Protection: ± 110 Vpk, continuous.
Range of Frequency: 40.000 00 Hz ~ 1000.000 kHz
Accuracy of Frequency: 20 ppm
Range of Phase: 0.000 0 ° ~ 359.999 9 °
Accuracy of Phase: 0.02 °@(40 Hz≤F≤200 Hz)
≤F≤1 kHz)
AC/DC Electric Energy / Power measurement
Measuring Range: Combination of voltage range and current range.
Accuracy: voltage accuracy + current accuracy.
Input of Standard Electric Energy Pulse: Directly receive the electric energy pulse of the checked meters.
Output of Standard Electric Energy Pulse:
The amplitude of the signal is 5 V, TTL Level.
The full-scale value of high frequency power corresponds to 60 kHz
The full-scale value of low frequency power corresponds to 6 Hz
Electric Energy Measurement Error: Automatically display with 0.0001%.
Specification
AC/DC Voltage Measurement
Range | Resolution | Accuracy (k = 2) ppm*RD +ppm*RG or V | Temperature Coefficient (/℃) | ||
DC | 40 Hz≤F≤200 Hz | 200Hz≤F≤1 kHz | |||
100 mV | 0.1 μV | 50 + 5 μ | 120 + 80 | 200 + 100 | 3 + 0.5 μ |
300 mV | 0.1 μV | 30 + 6 μ | 120 + 80 | 200 + 100 | 3 + 0.6 μ |
1 V | 1 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 1 |
3 V | 1 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 1 |
10 V | 10 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 0.5 |
30 V | 10 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 1 |
100 V | 100 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 1 |
300 V | 100 μV | 30 + 20 | 120 + 80 | 200 + 100 | 3.5 + 1.5 |
1000 V | 1 mV | 30 + 20 | 120 + 80 | 200 + 100 | 3.5 + 1.5 |
l Measuring Range of Voltage: ± (10 mV ~ 1150 V)
Range Switching: Automatic or Manual
Input Resistance: 1 MΩ
Range | Resolution | Accuracy (k = 2) ppm*RD +ppm*RG | Temperature Coefficient (/℃) | ||
DC | 40 Hz≤F≤200 Hz | 200Hz≤F≤1 kHz | |||
100 μA | 0.1 nA | 50 + 5 nA | 200 + 100 | 250 + 150 | 3 + 0.5 n |
300 μA | 0.1 nA | 50 + 15 nA | 200 + 100 | 250 + 150 | 3 + 1.5 n |
1 mA | 1 nA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
3 mA | 1 nA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
10 mA | 10 nA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
30 mA | 10 nA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
100 mA | 100 nA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
300 mA | 100 nA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
1 A | 1 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
3 A | 1 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
10 A | 10 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
30 A | 10 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
100 A | 100 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
200 A① | 100 μA | 30 + 20 | 200 + 100 | 250 + 150 | 3.5 + 1.5 |
500 A① | 100 μA | 30 + 20 | 200 + 100 | 250 + 150 | 3.5 + 1.5 |
Note: ①: Range of 200 A and 500 A are optional, which needs to be equipped with current sensor such as TH5450-500 A to realize range expansion
Measuring Range: ± (10 μA ~ 120 A)
Range Switching: Automatic or Manual
Direct input
DC Small signal voltage measurement DCIU (Indirect access)
Range | Resolution | Accuracy (k = 2) ppm*RD +ppm*RG | Temperature Coefficient (/℃) | ||
DC | 40 Hz≤F≤200 Hz | 200Hz≤F≤1 kHz | |||
1 mV | 1 nV | 40 + 0.2 μV | —— | —— | 4 + 0.02 μ |
10 mV | 10 nV | 40 + 0.5 μV | 120 + 80 | 200 + 100 | 3 + 0.05 μ |
100 mV | 100 nV | 40 + 10 | 80 + 40 | 100 + 50 | 3 + 0.5 |
1 V | 1 μV | 40 + 10 | 80 + 40 | 100 + 50 | 2 + 0.5 |
10 V | 10 μV | 40 + 10 | 80 + 40 | 100 + 50 | 2 + 0.5 |
Measuring Range: ± (0.1mV ~ 11 V)
Input Resistance: >1 GΩ
Input Protection: ± 110 Vpk, continuous.
Range of Frequency: 40.000 00 Hz ~ 1000.000 kHz
Accuracy of Frequency: 20 ppm
Range of Phase: 0.000 0 ° ~ 359.999 9 °
Accuracy of Phase: 0.02 °@(40 Hz≤F≤200 Hz)
≤F≤1 kHz)
AC/DC Electric Energy / Power measurement
Measuring Range: Combination of voltage range and current range.
Accuracy: voltage accuracy + current accuracy.
Input of Standard Electric Energy Pulse: Directly receive the electric energy pulse of the checked meters.
Output of Standard Electric Energy Pulse:
The amplitude of the signal is 5 V, TTL Level.
The full-scale value of high frequency power corresponds to 60 kHz
The full-scale value of low frequency power corresponds to 6 Hz
Electric Energy Measurement Error: Automatically display with 0.0001%.