TM2000 / 2030 Precision Current Source

The TM2000 / 2030 is a series of precision excitation current sources with high precision, high resolution regulation, high stability and other characteristics. It can be used as excitation power supply of field generator such as Helmholtz coil and solenoid to generate standard magnetic field. It’s applied for the influence test of materials, devices and instruments or the calibration of magnetic sensors and other applications.
Availability:
Quantity:
  • TM2000 / 2030

  • TUNKIA

TM2000 / 2030 Precision Current Source

TM2000 / 2030 Precision Current Source

*The drawing is for reference only. There may be some differences according to different technical specifications

1. Summary

The TM2000 / 2030 is a series of precision excitation current sources with high precision, high resolution regulation, high stability and other characteristics. It can be used as excitation power supply of field generator such as Helmholtz coil and solenoid to generate standard magnetic field. It’s applied for the influence test of materials, devices and instruments or the calibration of magnetic sensors and other applications.


2. Features

Single or three channel output are optional.

Current output can produce a magnetic field (±).

Optimal short-term stability is 20ppm/min.

Open loop or closed loop two test modes.

Open loop can set current or magnetic field value output.

Closed loop can realize feedback measurement with magnetic sensor.

Support a variety of value setting and adjustment.

Support cascaded expansion of current or power output.

Open circuit, overload, and overheating protection.




3. Specifications

3.1 Constant Current Output

Specifications

TM20X0-5A-30V

TM20X0-25A-12V

TM20X0-25A-24V

Current range

1A, 5A

1A, 5A, 25A

1A, 5A, 25A

Short-term stability ( ppm/min )

Class 0.05

100

100

100

Class 0.02

50

50

50

Class 0.01

30

30

30

Class 0.005

20

20

20

Accuracy

±(ppm of reading + ppm of range)

Class 0.05

300 + 200

300 + 200

300 + 200

Class 0.02

120 + 80

120 + 80

120 + 80

Class 0.01

60 + 40

60 + 40

60 + 40

Class 0.005

30 + 20

30 + 20

30 + 20

Maximum load voltage of

single channel

30 V

12 V

24 V

Maximum load Power of single channel

150W

300W

600W

Note: (ppm = parts per million) (e.g., 10ppm = 0.001%)

The above is the compliance voltage of the base model , and the maximum compliance voltage after cascade is proportional to the maximum power.



3.2 Magnetic Field Measurement(Optional)

Fluxgate





Range

100μT, 1mT, 2mT

Measuring range

1μT~2mT

Display digits

5-dights display

Probe

Single or triaxial fluxgate probes

Accuracy

±0.5%

Teslameter




Range

30mT, 300 mT, 3000 mT

Measuring range

2 mT~3600 mT

Display digits

5-dights display

Accuracy

0.5%


4. Ordering Information

TM2000 / 2030 Precision Current Source



Previous: 
Next: 

RELATED PRODUCTS

SIGN UP FOR NEWSLETTER

Get latest deals, newsletter and technical information

INFORMATION

CONTACT US

 WhatsApp:+86-18773180215
     Wechat: 18773180215
 Phone: +86-18773180215
 E-mail: global@tunkia.com
E-mail: Eugene@tunkia.com
 Address: NO.16 Panpan Road, Changsha
Copyright  2021 TUNKIA Co., Ltd. All Rights Reserved. Supported by leadong.comSitemap