1. Summary
TD1230 is designed based on TUNKIA's latest Integrated Precision Measurement Architecture (IPMA) (1). It is suitable for the calibration of Class 1, Class 2, and lower accuracy AC/DC withstand voltage testers. It can also be used to calibrate the withstand voltage test section of electrical safety multifunction testers and insulation withstand voltage testers.
2. Features
Measurement Capability
Accuracy class: 0.5 / 0.2 (optional).
AC/DC voltage range: 100 V to 5 kV / 10 kV / 15 kV / 30 kV.
AC/DC current range: 0.1 mA to 500 mA.
Measurement of AC distortion, DC ripple, and voltage duration.
Built-in adjustable resistive load with a minimum resolution of 0.1 kΩ for convenient calibration of leakage current and alarm current.
Other Features
Large color LCD touchscreen with rear-panel connections.
Built-in oscilloscope and measurement data statistics and analysis functions.
Connection guidance for high-voltage and leakage current measurements.
Compact and lightweight design for portable field applications.
Optional high-power load box (TD1240) for output capacity calibration.
3. Specifications
3.1 AC/DC Voltage Measurement
Range | Resolution | Measurement Uncertainty (k=2),( ppm*RD+ppm*RG )[1] | |
Class 0.5 | Class 0.2 | ||
5 kV | 0.1 V | 4000 + 1000 | 1800 + 200 |
10 kV[2] | 1 V | 4000 + 1000 | 1800 + 200 |
15 kV[2] | 1 V | 4000 + 1000 | 1800 + 200 |
30 kV[2] | 1 V | 4000 + 1000 | 1800 + 200 |
Note [1] : RD is the reading value, RG is the range value. Note [2]: 10 kV, 15 kV, 30 kV ranges are optional | |||
Measuring range: 100 V to 30 kV, 5-bit display, manual/automatic range switching
Input impedance: 600 MΩ, true RMS response for AC, average response for DC
2.2 AC/DC Current Measurement
Range | Resolution | Measurement Uncertainty (k=2),( ppm*RD+ppm*RG )[1] | |
Class 0.5 | Class 0.2 | ||
2 mA | 0.1 μA | 4000 + 1000 | 1800 + 200 |
20 mA | 1 μA | 4000 + 1000 | 1800 + 200 |
200 mA | 10 μA | 4000 + 1000 | 1800 + 200 |
400 mA | 10 μA | 4000 + 1000 | 1800 + 200 |
Measuring range: 0.1 mA~500 mA, 5-digit display, manual/automatic switching range, margin: 125%*RG.
Built-in adjustable load resistance: 2.5 kΩ to 8 MΩ; minimum resolution: 0.1 kΩ.
True RMS response for AC, average response for DC
2.3 Other Functions
Test Items | Range | Resolution | Measurement Uncertainty (k=2) |
AC Frequency | 45 Hz~65 Hz | 0.1 Hz | 0.2% |
AC Voltage Distortion | 0.5%~10% | 0.01% | 0.5% |
DC Voltage Ripple | 0.1%~30% | 0.01% | 0.2% |
Voltage Duration Time | 0~20 s | 0.001 s | 0.2 s |
20 s~1000 s | 1% |
6. General Specifications
| Warm-Up Time | 30 min |
Maximum Power Consumption | 80 VA |
Temperature Performance | Working temperature: 0 °C~45 °C; Storage temperature: -20 °C~70 °C |
Humidity Performance | Working humidity: < 80% @ 30 °C, < 70% @ 40 °C, < 40% @ 50 °C Storage humidity: (20%~80%) R·H, non-condensing |
Altitude | < 3000 m |
Weight | Approx. 6 kg |
Communication Interface | LAN, USB |
Dimensions | 212 mm( W ) × 350 mm( D ) × 132 mm( H ) (excluding handles and feet) |
|
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment