TS2830 Reverse Bend Test and Analysis System for Electrical Steel Sheets
(Original model: TS1800)
*Above picture is only for reference, subject to actual delivery
TS2830 is a dedicated testing device for performing bending and reverse flat bending tests on electrical steel sheets. It is equipped with a specialized clamping fixture for sheet specimens, and is designed to evaluate the ability of individual electrical steel sheets to withstand plastic deformation under repeated bending, as well as to reveal any resulting defects. Reference standard: YB/T 4731-2019 Reverse bend test method for electrical steel strip and sheet (Tunkia participated in drafting).
Quick clamping of electrical steel sheet.
The device can be easily fixed on the desktop by tightening screws and a vise at the bottom.
The bending radius of the lower clamp is 5 mm, and the bending radius is not less than ±90°.
The upper spring can exert stable tension on electrical steel sheet.
The number of bends can be displayed
Bending Radius | 5 mm |
Bending Angle | ± 90° |
Counting Range | 0 ~ 9999 |
Specimens Size | 30 mm × ( 280 ~ 320 ) mm (width × length) |
Power Supply | AC ( 220 ± 22 ) V, ( 50 ± 2 ) Hz |
Temperature Performance | Working temperature: 0°C ~ 40°C; Storage temperature: -20°C ~ 70°C |
Humidity Performance | Working humidity: 40% ~ 80% R·H,No condensation Store humidity: < 95% R·H,No condensation |
Dimension | 255 mm(W) ×180 mm(D) × 565 mm(H) |
Step | Position | Description |
1 | | Insert the fixing pin to keep the device vertical |
2 | | Insert the silicon steel sheet to be tested and keep it in position |
3 | | Turn the upper clamping handle to clamp the upper part of the silicon steel sheet |
4 | | Turn the spring clamping handle to compress the spring |
5 | | Turn the lower clamping handle to clamp the lower part of the silicon steel sheet |
6 | | Turn the clamping handle in the opposite direction, the spring is released, and the silicon steel sheet is tightened. |
7 | | Press the RESET button below the digital counter display to reset the digital counter to zero. |
8 | | Pull out the fixing pin and prepare to start the repeated bending test. |
9 | | Hold the upper handle and start the repeated bending test. When turning ±90 degrees, the digital counter will automatically count. |
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment