1. Summary
TD2650 is a high-precision AC/DC standard meter with a wide measurement range. It is capable of measuring up to 50 kV DC high voltage and 10 kV AC high voltage, and is available in two accuracy classes: 0.02 and 0.05. The instrument is suitable for verifying or calibrating AC/DC high-voltage sources, AC/DC high-voltage meters, dielectric withstand test devices, high-voltage probes, and similar equipment.
2. Features
DC Voltage Standard Meter: 100 V–5 kV / 10 kV / 20 kV / 50 kV.
AC Voltage Standard Meter: 100 V–5 kV / 10 kV
Stability testing and data statistical analysis functions
Reliable electrical protection, including short-circuit and overload protection
Large LCD touch screen for intuitive value display and convenient operation
Equipped with an RS232 interface for easy connection with dedicated software to build an automated testing system
3. Application
☆ Calibration for AC or DC High Voltage Source / Meter |
|
• DCV measurement: 100 V ~ 5 kV / 10 kV / 20 kV / 50 kV • ACV measurement: 100 V ~ 5 kV / 10 kV • Apply for calibrating DC high voltage source or meter, AC high voltage source or meter, high voltage probe, etc. |
4. Characteristics
☆ Data Statistical Analysis Function | |
| |
S/N | Function |
| Statistical analysis of data: Maximum value (Max), Minimum value (Min), Peak-to-Peak value (P-P) Average value (Avg), Standard variance (Std), Stability (Stab), etc. |
5. Specifications
5.1 DCV Measurement
Model | Range (kV) | Accuracy ±(ppm of reading + ppm of range) [1] | Temperature Coefficient (±ppm/K) | ||
Class 0.05 | Class 0.02 | Class 0.05 | Class 0.02 | ||
TD2650-5kV-xxx | 1, 2, 5 | 300 + 200 | 120 + 80 | < 15 | < 10 |
TD2650-10kV-xxx | 1, 2, 5, 10 | 300 + 200 | 120 + 80 | < 15 | < 10 |
TD2650-20kV-xxx | 2, 5, 10, 20 | 300 + 200 | 120 + 80 | < 15 | < 10 |
TD2650-50kV-xxx | 5, 10, 20, 50 | 300 + 200 | - | < 15 | - |
Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%). | |||||
• Measurement range :(10% ~ 100%)*range, 6 digits display,
• Manual or automatic range switch
5.2 ACV Measurement
Model | Range (kV) | Accuracy ±(ppm of reading + ppm of range) | Temperature Coefficient (±ppm/K) | ||
Class 0.05 | Class 0.02 | 0.05 | 0.02 | ||
TD2650-xxx-5kV | 2, 5 | 300 + 200 | 120 + 80 | < 25 | < 15 |
TD2650-xxx-10kV | 2, 5, 10 | 300 + 200 | 120 + 80 | < 25 | < 15 |
• measurement range :(10% ~ 100%)*RG, 6 digits display, manual/automatic range switch
• Frequency: 45.000 0 Hz ~ 65.000 0 Hz, uncertainty: 0.01 Hz
• Max. voltage 25 kV is customable.
6. General Specifications
Power Supply | AC (220 ± 22) V,(50 ± 2) Hz |
Temperature | Working temperature:0℃~45℃; Storage temperature:-20℃~70℃ |
Humidity | Working humidity:< 80% @ 30℃,< 70% @ 40℃,< 40% @ 50℃ Storage humidity:(20%~80%) R·H,non-condensing |
Interface | RS232 interface |
| Dimension | 520 mm(W) × 450 mm(D) × 205 mm(H) |
*Note: The sizes of different models may vary slightly. Here we take TD2650-500-10kV-10kV as an example. |
7. Ordering Information
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment