TD1470 Single and Double Arm Bridge Calibration Device

    TD1470 is an instrument that can accurately simulate a wide range of DC standard resistance. It can realize the verification and calibration of single-arm bridges, double-arm bridges, high-precision DC resistance meters and digital micro-ohmmeters.
Availability:
Quantity:
  • TD1470

  • TUNKIA

TD1470 Single and Double Arm Bridge Calibration Device


TD1470 Single and Double Arm Bridge Calibration Device


1. Summary


TD1470 is an instrument that can accurately simulate a wide range of DC standard resistance. It can realize the verification and calibration of single-arm bridges, double-arm bridges, high-precision DC resistance meters and digital micro-ohmmeters.



2. Features


DC current measurement range: 1 mA ~ 11 A

Resistance simulation range: 100 μΩ ~ 11 MΩ

Best measurement uncertainty of resistance: 100 ppm

Two-wire and four-wire resistance modes

RS232 interface

Large-size LCD touch screen


3. Application


Calibrate single-arm bridge with class 0.05 and below

Calibrate DC resistance meters of class 0.05 and below

Calibrate temperature measurement bridges of class 0.05 and below


4. Characteristics

Analog resistance output

TD1470 Single and Double Arm Bridge Calibration Device

    • The four-wire wiring method can effectively eliminate the measurement error caused by the resistance of the test wire, and the measurement accuracy is higher;

    • Compared with physical resistor boxes, analog resistors have the following advantages:

The resistance value is continuously adjustable, with better adjustment fineness and better sensitivity;

There is no influence from switching error, residual resistance error, contact resistance error, etc., and the measurement data is accurate and reliable;

When the physical resistor is overloaded, it will cause the resistance to change value, and in severe cases, it may even cause the resistor to be damaged. TD1470 has current overload capability and impact resistance, and has been completely tested for electrostatic protection and electromagnetic compatibility. It has good protection performance and high reliability.

TD1470 has a self-calibration function to ensure long-term accuracy and stability of the value.


Easy to operate

TD1470 Single and Double Arm Bridge Calibration Device

Large-size LCD screen: full color, high screen brightness, clear picture quality, and supports touch operation.

Digit buttons: A variety of input methods such as fixed-point output, knob output, and step adjustment can be realized, and the operation is convenient and fast.

Front panel wiring: It is convenient for users to replace the current and voltage test wires when checking the meter.



Multiple output methods

TD1470 Single and Double Arm Bridge Calibration Device

The instrument has a "fixed point output" mode, through the digital button of the control panel or click the touch screen, directly set the required output value, the instrument will automatically switch to the best range output.


Multiple output methods

TD1470 Single and Double Arm Bridge Calibration Device

When calibrating a resistance tester, it is usually necessary to select the calibration point according to the proportion of each range of the meter to be checked.

The user can easily select the calibration point of the inspected meter through the "Percentage Output Button" on the console of this instrument or the "Percentage Check Point" on the touch screen.

TD1470 Single and Double Arm Bridge Calibration Device

The operating area is equipped with a "Number Knob" that increases or decreases the output by rotating it clockwise or counterclockwise.


Statistical analysis of test data

TD1470 Single and Double Arm Bridge Calibration Device

S/N

Function

Integration period: The integration period can be set in the range of 0.5 s ~ 20 s (default 1 s);

Statistical analysis: Real-time recording and statistical test data, including current Max, Min, Avg, Span, Stab, S. dev, Time, test times N, etc.


Rich communication interface and special test software

TD1470 Single and Double Arm Bridge Calibration Device

RS232, USB, LAN interfaces.

5. Specifications

5.1 Low Value Resistor Simulation

Range


Fineness Adjustment


Measurement Uncertainty (k=2, ppm*RD+μΩ)[1]

Input Current[2]


24h

( 23 ± 1 )°C

90days

( 23 ± 5 )°C

1year

( 23 ± 5 )°C

1 mΩ

10 nΩ

50 + 0.5

75 + 0.5

100 + 0.5

1 A~10 A

10 mΩ

100 nΩ

50 + 1

75 + 1

100 + 1

0.5 A~5 A

100 mΩ

1 μΩ

50 + 3

75 + 3

100 + 3

0.3 A~5 A

1 Ω

10 μΩ

50 + 15

75 + 15

100 + 15

0.05 A~5 A

10 Ω

100 μΩ

50 + 150

75 + 150

100 + 150

10 mA~300 mA

100 Ω

1 mΩ

50 + 1500

75 + 1500

100 + 1500

1 mA~30 mA

Note [1]: RD is the reading value, the same below;;

[2] When the current and resistance exceed the range, the resistance value will bring additional error

• Resistance Simulation Range: (10%~110%)*RG,6-digit display

• Wiring Method: 4-wire

• Current Input Range: 1 mA ~ 11 A

• Range Switching: manual/automatic shifting


5.2 Medium and High Value Resistance Simulation


Range


Fineness Adjustment


Measurement Uncertainty ( k=2, ppm*RD+Ω )

Input Current


24h

(23±1)°C

90days

(23±5)°C

1year

(23±5)°C

100 Ω

1 mΩ

50 + 0.002

75 + 0.002

100 + 0.002

1 mA~80 mA

1 kΩ

10 mΩ

45 + 0.01

50 + 0.01

90 + 0.01

1 mA~12 mA

10 kΩ

100 mΩ

45 + 0.1

50 + 0.1

90 + 0.1

100 μA~2 mA

100 kΩ

1 Ω

45 + 1.0

50 + 1.0

90 + 1.0

10 μA~0.2 mA

1 MΩ

10 Ω

50 + 10

75 + 10

100 + 10

1 μA~20 μA

10 MΩ

100 Ω

100 + 100

150 + 100

200 + 100

250 nA~2 μA

• Resistance Simulation Range: (10%~110%)*RG,6-digit display

• Wiring Method: 4-wires /2-wires

• Current Input Range: 250 nA ~ 80 mA

• Range Switching: manual/automatic shifting


6. General Specifications

Power Supply

AC ( 220 ± 22 ) V( 50 ± 2 ) Hz

Warm-up Time

Not less than 1 hour, and the warm-up time again after shutting down in the middle should not be less than 2 times the shutdown time;

Maximum Power Consumption

150 VA

Temperature Performance

Working temperature: 18 °C ~ 28 °C

Storage temperature: -10 °C ~ 50 °C

Humidity Performance

Working humidity:  ( 20 % ~ 80% ) R·H, non-condensing;

Storage humidity:< 85% R·H, no condensation;

Altitude

< 3000 m

Weight

Approx 10 kg

Communication Interface

RS 232、USB、IP

Dimension


400 mm ( W ) x 265 mm ( D ) x 190 mm ( H ) (without feet and handles)

TD1470 Single and Double Arm Bridge Calibration Device


7. Accessories List

NO.

Picture

      Name

     Specification

      Quantity

Remark

1

High Value Resistance Test Leads

6 in 1,

0.8m-Φ4 gun stock -Φ4 gun stock

1

Standard Accessory

2

Universal Serial Cable

1.8m / USB to RS232(DB9 socket)

1

Standard Accessory

3

Power Cable

AC 220V、10A

1

Standard Accessory

4

Packaging Box

Aluminum Alloy Box

1

Standard Accessory


NO.

Picture

     Name

      Specification

    Quantity

Remark

1

Calibration Software

Card USB

1

Optional Accessory

2

Packaging Box

Pelican

1

Optional Accessory

Note: The above optional accessory need to be purchased separately and specified in the order contract.


8. Relevant Product

• DC current measurement range: 20 mA ~ 22 A

• Best current measurement uncertainty: 40 ppm

• Resistance simulation range: 0 Ω ~ 1.1 GΩ

• Best measurement uncertainty of resistance: 20 ppm


TH0360 High Precision DC Resistance Meter Calibrator

• DC current measurement range: 1 mA ~ 11 A

• Resistance simulation range: 100 μΩ ~ 11 MΩ

• Best measurement uncertainty of resistance: 50 ppm


TH0350 DC Bridges Calibrator

Previous: 
Next: 

SIGN UP FOR NEWSLETTER

Get latest deals, newsletter and technical information

INFORMATION

CONTACT US

 WhatsApp:+86-18773180215
     Wechat: 18773180215
 Phone: +86-18773180215
 E-mail: global@tunkia.com
E-mail: Eugene@tunkia.com
 Address: NO.16 Panpan Road, Changsha
Copyright  2021 TUNKIA Co., Ltd. All Rights Reserved. Supported by leadong.comSitemap