1. Summary
TH0360 is an instrument that accurately simulates a wide range of DC standard resistances, enabling single-arm bridges, double-arm bridges, high-precision DC resistance meters, and digital microohmmeters.
2. Features
• DC current measurement range: 20 mA~ 22 A
• Optimal measurement uncertainty for current: 40 ppm
• Resistance analog range: 0 Ω ~ 1.1 GΩ
• Optimal measurement uncertainty for resistance: 20 ppm
• Two-wire and four-wire resistance modes
• RS232 interface
• LCD touch screen
3. Applications
• Calibrate a class 0.01 and below single-arm bridge
• Calibrate DC resistance meters of class 0.01 and below
• Calibrate a 6-digit half-digit multimeter
• Calibrate DC current sources below class 0.01
• Calibrate temperature measurement bridges of 0.01 and below
4. Characteristics
☆ Analog resistance output |
• Higher measurement accuracy with the four-wire wiring method. • Self-calibration. |
☆ Multiple output methods |
• The instrument has a "fixed point output" mode, through the digital button of the Keypad or click the touch screen, directly set the required output value, the instrument will automatically switch to the best range output. |
☆ Multiple output methods |
• When calibrating a resistance tester, it is usually necessary to select the calibration point according to the proportion of each range of the meter to be checked. • The user can easily select the calibration point of the inspected meter through the "Percentage Output Button" on the console of this instrument or the "Percentage Check Point" on the touch screen. |
• The operating area is equipped with a "rotary encoder" that increases or decreases the output by rotating it clockwise or counterclockwise. |
☆ Statistical analysis of test data | ||
S/N | Function | |
Integration period: The integration period can be set in the range of 0.5 s ~ 20 s (default 1 s), which is convenient for users to set sampling data according to actual needs; | ||
Statistical analysis: real-time recording and statistical test data, including current average Avg, mean square deviation S.dev, peak-to-peak P-P, and maximum value Max, minimum min, test number N, time time, etc., easy for users to read intuitively; |
☆ Rich communication interface and special test software |
• RS232, USB, LAN interfaces. |
1. Summary
TH0360 is an instrument that accurately simulates a wide range of DC standard resistances, enabling single-arm bridges, double-arm bridges, high-precision DC resistance meters, and digital microohmmeters.
2. Features
• DC current measurement range: 20 mA~ 22 A
• Optimal measurement uncertainty for current: 40 ppm
• Resistance analog range: 0 Ω ~ 1.1 GΩ
• Optimal measurement uncertainty for resistance: 20 ppm
• Two-wire and four-wire resistance modes
• RS232 interface
• LCD touch screen
3. Applications
• Calibrate a class 0.01 and below single-arm bridge
• Calibrate DC resistance meters of class 0.01 and below
• Calibrate a 6-digit half-digit multimeter
• Calibrate DC current sources below class 0.01
• Calibrate temperature measurement bridges of 0.01 and below
4. Characteristics
☆ Analog resistance output |
• Higher measurement accuracy with the four-wire wiring method. • Self-calibration. |
☆ Multiple output methods |
• The instrument has a "fixed point output" mode, through the digital button of the Keypad or click the touch screen, directly set the required output value, the instrument will automatically switch to the best range output. |
☆ Multiple output methods |
• When calibrating a resistance tester, it is usually necessary to select the calibration point according to the proportion of each range of the meter to be checked. • The user can easily select the calibration point of the inspected meter through the "Percentage Output Button" on the console of this instrument or the "Percentage Check Point" on the touch screen. |
• The operating area is equipped with a "rotary encoder" that increases or decreases the output by rotating it clockwise or counterclockwise. |
☆ Statistical analysis of test data | ||
S/N | Function | |
Integration period: The integration period can be set in the range of 0.5 s ~ 20 s (default 1 s), which is convenient for users to set sampling data according to actual needs; | ||
Statistical analysis: real-time recording and statistical test data, including current average Avg, mean square deviation S.dev, peak-to-peak P-P, and maximum value Max, minimum min, test number N, time time, etc., easy for users to read intuitively; |
☆ Rich communication interface and special test software |
• RS232, USB, LAN interfaces. |
5. Specifications
5.1 DC current measurement
Current range | Resolution | Measurement uncertainty (k=2, ppm*RD + ppm*RG) [1]. | Temperature coefficient (ppm*RD/°C) @ (15~30)°C | ||
24 hours ( 23 ± 1 ) °C | 90 days ( 23 ± 5 ) °C | 1 year ( 23 ± 5 ) °C | |||
200 mA | 10 nA | 18 + 6 | 24 + 8 | 30 + 10 | 0.5 |
1 A | 100 nA | 18 + 6 | 24 + 8 | 30 + 10 | 0.5 |
5 A | 1 μA | 18 + 6 | 24 + 8 | 30 + 10 | 0.5 |
20 A | 10 μA | 18 + 6 | 24 + 8 | 30 + 10 | 0.5 |
Note [1]: RD is the reading value and RG is the range value |
• Measuring range: 20 m A ~ 22 A
• Display digits: 6 digits display
• Range switching: manual/automatic switching
5.2 Low value resistance simulation
Test the current | Resistance range | Adjust fineness | Measurement uncertainty ( k=2 ) |
5 A ≤ I ≤ 22 A | 0 ≤ R ≤ 50 mΩ | 10 pH | 25 ppm*RD + KI[2]*4.0 nΩ |
50 μΩ < R ≤ 500 μΩ | 0.1 nΩ | 25 ppm*RD + KI*5.0 nΩ | |
500 μΩ < R ≤ 5 mΩ | 1 nΩ | 25 ppm*RD + KI*10 nΩ | |
5 mΩ < R ≤ 50 mΩ | 10 nΩ | 25 ppm*RD | |
50 mΩ < R ≤ 500 mΩ | 0.1 mΩ | 25 ppm*RD | |
1 A ≤ I ≤ 5.5 A | 0 ≤ R ≤ 200 mΩ | 0.1 nΩ | 25 ppm*RD + KI*16 nΩ |
200 μΩ < R ≤ 2 mΩ | 1 nΩ | 25 ppm*RD + KI*20 nΩ | |
2 mΩ < R ≤ 20 mΩ | 10 nΩ | 25 ppm*RD + KI*50 nΩ | |
20 mΩ < R ≤ 200 mΩ | 100 nΩ | 25 ppm*RD | |
200 mΩ < R ≤ 2 Ω | 1 mΩ | 25 ppm*RD | |
0.2 A ≤ I ≤ 1.1 A | 0 ≤ R ≤ 1 mΩ | 1 nΩ | 25 ppm*RD + KI*80 nΩ |
1 mΩ < R ≤ 10 mΩ | 10 nΩ | 25 ppm*RD + KI*100 nΩ | |
10 mΩ < R ≤ 100 mΩ | 100 nΩ | 25 ppm*RD + KI*200 nΩ | |
100 mΩ < R ≤ 1 Ω | 1 mΩ | 25 ppm*RD | |
1 Ω < R ≤ 10 Ω | 10 mΩ | 25 ppm*RD | |
50 mA ≤ I ≤ 220 mA | 0 ≤ R ≤ 5 mΩ | 1 nΩ | 25 ppm*RD + KI*0.4 μΩ |
5 mΩ < R ≤ 50 mΩ | 10 nΩ | 25 ppm*RD + KI*0.5 μΩ | |
50 mΩ < R ≤ 500 mΩ | 100 nΩ | 25 ppm*RD + KI*1.0 μΩ | |
500 mΩ < R ≤ 5 Ω | 1 mΩ | 25 ppm*RD | |
5 Ω < R ≤ 50 Ω | 10 mΩ | 25 ppm*RD | |
Note [2]: KI is the current coefficient, KI = RG/I, where RG is the current range value and I is the input current value. |
• Wiring mode: four-wire system
• Current input range: 50 mA ~ 22 A
• Range switching: manual/automatic switching
5.3 Medium and high value resistor simulation
Range | Adjust fineness | Measurement uncertainty ( k=2, ppm*RD+Ω ) | Input current | ||
24 hours (23±1)°C | 90 days (23±5)°C | 1 year (23±5)°C | |||
10 Ω | 100 mΩ | 20 + 0.001 | 32 + 0.001 | 40 + 0.001 | 1 mA ~ 150 mA |
30 Ω | 150 mΩ | 15 + 0.002 | 24 + 0.002 | 30 + 0.002 | 1 mA ~ 150 mA |
100 Ω | 500 mΩ | 12 + 0.002 | 20 + 0.002 | 25 + 0.002 | 1 mA ~ 80 mA |
300 Ω | 1.5 mΩ | 12 + 0.003 | 20 + 0.003 | 25 + 0.003 | 1 mA ~ 60 mA |
1 kΩ | 5 mΩ | 12 + 0.01 | 18 + 0.01 | 20 + 0.01 | 1 mA ~ 12 mA |
3 kΩ | 15 mΩ | 12 + 0.03 | 18 + 0.03 | 20 + 0.03 | 0.1 mA ~ 4 mA |
10 kΩ | 50 mΩ | 12 + 0.1 | 18 + 0.1 | 20 + 0.1 | 0.1 mA ~ 2 mA |
30 kΩ | 150 mΩ | 12 + 0.3 | 18 + 0.3 | 20 + 0.3 | 10 μA ~ 400 μA |
100 kΩ | 500 mΩ | 15 + 1 | 24 + 1 | 30 + 1 | 10 μA ~ 200 μA |
300 kΩ | 1.5 Ω | 17 + 3 | 28 + 3 | 35 + 3 | 1 μA ~ 40 μA |
1 MΩ | 5 Ω | 17 + 10 | 28 + 10 | 35 + 10 | 1 μA ~ 20 μA |
3 MΩ | 15 Ω | 30 + 30 | 48 + 30 | 60 + 30 | 0.25 μA ~ 4 μA |
10 MΩ | 50Ω | 65 + 100 | 110 + 100 | 130 + 100 | 0.25 μA ~ 2 μA |
30 MH | 150 Ω | 125 + 2.5k | 150 + 2.5k | 200 + 2.5k | 25 nA ~ 400 nA |
100 MΩ | 1 kΩ | 250 + 3k | 400 + 3k | 500 + 3k | 25 nA ~ 200 nA |
300 MΩ | 1.5 kΩ | 1200 + 0.1M | 1500 + 0.1M | 2000 + 0.1M | 2.5 nA ~ 70 nA |
1 GΩ | 5 kΩ | 6000 + 0.5M | 7000 + 0.5M | 8000 + 0.5M | 1 nA ~ 20 nA |
• Resistance analog range: 1 Ω ~ 1.1 G Ω
• Wiring mode: four-wire system or two-wire system
• Current input range: 1 nA ~ 150 mA
• Range switching: manual/automatic switching
6. General Specifications
Power supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz; |
Warm-up time | Not less than 1 hour, and the warm-up time again after shutting down in the middle should not be less than 2 times the shutdown time; |
Maximum power consumption | 150 VA |
Temperature performance | Working temperature: 18 °C ~ 28 °C; Storage temperature: -10 °C ~ 50 °C; |
Humidity performance | Operating humidity:( 20 % ~ 80% ) R H, no condensation; Storage humidity: < 85% R·H, no condensation; |
Altitude | < 3000 m |
Quality | Approx. 13 kg |
Interfaces | RS 232、USB、IP |
Size | 400 mm ( W )× 300 mm ( D ) × 190 mm ( H ) ( without feet and handles). |
5. Specifications
5.1 DC current measurement
Current range | Resolution | Measurement uncertainty (k=2, ppm*RD + ppm*RG) [1]. | Temperature coefficient (ppm*RD/°C) @ (15~30)°C | ||
24 hours ( 23 ± 1 ) °C | 90 days ( 23 ± 5 ) °C | 1 year ( 23 ± 5 ) °C | |||
200 mA | 10 nA | 18 + 6 | 24 + 8 | 30 + 10 | 0.5 |
1 A | 100 nA | 18 + 6 | 24 + 8 | 30 + 10 | 0.5 |
5 A | 1 μA | 18 + 6 | 24 + 8 | 30 + 10 | 0.5 |
20 A | 10 μA | 18 + 6 | 24 + 8 | 30 + 10 | 0.5 |
Note [1]: RD is the reading value and RG is the range value |
• Measuring range: 20 m A ~ 22 A
• Display digits: 6 digits display
• Range switching: manual/automatic switching
5.2 Low value resistance simulation
Test the current | Resistance range | Adjust fineness | Measurement uncertainty ( k=2 ) |
5 A ≤ I ≤ 22 A | 0 ≤ R ≤ 50 mΩ | 10 pH | 25 ppm*RD + KI[2]*4.0 nΩ |
50 μΩ < R ≤ 500 μΩ | 0.1 nΩ | 25 ppm*RD + KI*5.0 nΩ | |
500 μΩ < R ≤ 5 mΩ | 1 nΩ | 25 ppm*RD + KI*10 nΩ | |
5 mΩ < R ≤ 50 mΩ | 10 nΩ | 25 ppm*RD | |
50 mΩ < R ≤ 500 mΩ | 0.1 mΩ | 25 ppm*RD | |
1 A ≤ I ≤ 5.5 A | 0 ≤ R ≤ 200 mΩ | 0.1 nΩ | 25 ppm*RD + KI*16 nΩ |
200 μΩ < R ≤ 2 mΩ | 1 nΩ | 25 ppm*RD + KI*20 nΩ | |
2 mΩ < R ≤ 20 mΩ | 10 nΩ | 25 ppm*RD + KI*50 nΩ | |
20 mΩ < R ≤ 200 mΩ | 100 nΩ | 25 ppm*RD | |
200 mΩ < R ≤ 2 Ω | 1 mΩ | 25 ppm*RD | |
0.2 A ≤ I ≤ 1.1 A | 0 ≤ R ≤ 1 mΩ | 1 nΩ | 25 ppm*RD + KI*80 nΩ |
1 mΩ < R ≤ 10 mΩ | 10 nΩ | 25 ppm*RD + KI*100 nΩ | |
10 mΩ < R ≤ 100 mΩ | 100 nΩ | 25 ppm*RD + KI*200 nΩ | |
100 mΩ < R ≤ 1 Ω | 1 mΩ | 25 ppm*RD | |
1 Ω < R ≤ 10 Ω | 10 mΩ | 25 ppm*RD | |
50 mA ≤ I ≤ 220 mA | 0 ≤ R ≤ 5 mΩ | 1 nΩ | 25 ppm*RD + KI*0.4 μΩ |
5 mΩ < R ≤ 50 mΩ | 10 nΩ | 25 ppm*RD + KI*0.5 μΩ | |
50 mΩ < R ≤ 500 mΩ | 100 nΩ | 25 ppm*RD + KI*1.0 μΩ | |
500 mΩ < R ≤ 5 Ω | 1 mΩ | 25 ppm*RD | |
5 Ω < R ≤ 50 Ω | 10 mΩ | 25 ppm*RD | |
Note [2]: KI is the current coefficient, KI = RG/I, where RG is the current range value and I is the input current value. |
• Wiring mode: four-wire system
• Current input range: 50 mA ~ 22 A
• Range switching: manual/automatic switching
5.3 Medium and high value resistor simulation
Range | Adjust fineness | Measurement uncertainty ( k=2, ppm*RD+Ω ) | Input current | ||
24 hours (23±1)°C | 90 days (23±5)°C | 1 year (23±5)°C | |||
10 Ω | 100 mΩ | 20 + 0.001 | 32 + 0.001 | 40 + 0.001 | 1 mA ~ 150 mA |
30 Ω | 150 mΩ | 15 + 0.002 | 24 + 0.002 | 30 + 0.002 | 1 mA ~ 150 mA |
100 Ω | 500 mΩ | 12 + 0.002 | 20 + 0.002 | 25 + 0.002 | 1 mA ~ 80 mA |
300 Ω | 1.5 mΩ | 12 + 0.003 | 20 + 0.003 | 25 + 0.003 | 1 mA ~ 60 mA |
1 kΩ | 5 mΩ | 12 + 0.01 | 18 + 0.01 | 20 + 0.01 | 1 mA ~ 12 mA |
3 kΩ | 15 mΩ | 12 + 0.03 | 18 + 0.03 | 20 + 0.03 | 0.1 mA ~ 4 mA |
10 kΩ | 50 mΩ | 12 + 0.1 | 18 + 0.1 | 20 + 0.1 | 0.1 mA ~ 2 mA |
30 kΩ | 150 mΩ | 12 + 0.3 | 18 + 0.3 | 20 + 0.3 | 10 μA ~ 400 μA |
100 kΩ | 500 mΩ | 15 + 1 | 24 + 1 | 30 + 1 | 10 μA ~ 200 μA |
300 kΩ | 1.5 Ω | 17 + 3 | 28 + 3 | 35 + 3 | 1 μA ~ 40 μA |
1 MΩ | 5 Ω | 17 + 10 | 28 + 10 | 35 + 10 | 1 μA ~ 20 μA |
3 MΩ | 15 Ω | 30 + 30 | 48 + 30 | 60 + 30 | 0.25 μA ~ 4 μA |
10 MΩ | 50Ω | 65 + 100 | 110 + 100 | 130 + 100 | 0.25 μA ~ 2 μA |
30 MH | 150 Ω | 125 + 2.5k | 150 + 2.5k | 200 + 2.5k | 25 nA ~ 400 nA |
100 MΩ | 1 kΩ | 250 + 3k | 400 + 3k | 500 + 3k | 25 nA ~ 200 nA |
300 MΩ | 1.5 kΩ | 1200 + 0.1M | 1500 + 0.1M | 2000 + 0.1M | 2.5 nA ~ 70 nA |
1 GΩ | 5 kΩ | 6000 + 0.5M | 7000 + 0.5M | 8000 + 0.5M | 1 nA ~ 20 nA |
• Resistance analog range: 1 Ω ~ 1.1 G Ω
• Wiring mode: four-wire system or two-wire system
• Current input range: 1 nA ~ 150 mA
• Range switching: manual/automatic switching
6. General Specifications
Power supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz; |
Warm-up time | Not less than 1 hour, and the warm-up time again after shutting down in the middle should not be less than 2 times the shutdown time; |
Maximum power consumption | 150 VA |
Temperature performance | Working temperature: 18 °C ~ 28 °C; Storage temperature: -10 °C ~ 50 °C; |
Humidity performance | Operating humidity:( 20 % ~ 80% ) R H, no condensation; Storage humidity: < 85% R·H, no condensation; |
Altitude | < 3000 m |
Quality | Approx. 13 kg |
Interfaces | RS 232、USB、IP |
Size | 400 mm ( W )× 300 mm ( D ) × 190 mm ( H ) ( without feet and handles). |