You are here: Home » Products » High Voltage Standard Instrument » TD2410 High Voltage Meters Calibration Device » TD2410 High Voltage Meters Calibration Device

TD2410 High Voltage Meters Calibration Device

TD2410 is a high precision and multi-function AC standard high voltage source with wide rang. It’s applied for calibrating AC high voltage meter, withstanding voltage testers, high voltage probe etc.
  • TD2410

  • TUNKIA

Features:
Availability:

TD2410 High Voltage Meters Calibration Device

TUNKIA TD2410 High Voltage Meters Calibration Device

1. Summary

TD2410 is a high precision and multi-function AC standard high voltage source with wide rang. It’s applied for calibrating AC high voltage meter, withstanding voltage testers, high voltage probe etc.


2. Features

AC voltage output: 1kV~35 kV.

Accuracy class: 0.02, 0.05.

Frequency: 45 Hz~65 Hz.

• 2nd~11th harmonic output (option).

• Short-circuit and overload protection.

• LCD touch screen

• High voltage switch for safety.

• RS232 interface.


3. Application

☆ Calibration for AC HV meter or Probe

TUNKIA TD2410 High Voltage Meters Calibration Device

TUNKIA TD2410 High Voltage Meters Calibration Device

• AC voltage ( ACV ) output: 1 kV~35 kV

• Apply to the calibration of alternating high voltage meter,high voltage probe etc.



4. Characteristics

☆ Security of Operation

TUNKIA TD2410 High Voltage Meters Calibration DeviceTUNKIA TD2410 High Voltage Meters Calibration Device

• High voltage output indicator which can warning high voltage danger

• Optical grating (option): Optical grating is controlled by software,if someone passing through the optical grating TD2410 will cut off the output immediately for safety.


☆ Multiple Output Mode

TUNKIA TD2410 High Voltage Meters Calibration Device

• “Direct output”mode,User can set output value by touch screen.

TUNKIA TD2410 High Voltage Meters Calibration Device

• Touch “Calibration point”of screen for “% setting”.


☆ Harmonic Function

TUNKIA TD2410 High Voltage Meters Calibration Device

S/N

Function

It can output 2nd~11th harmonic, and set harmonic amplitude.

Users can select the harmonics and set up harmonic parameter

Harmonic interface for output.

5. Specifications

Range

Resolution

Stability

(%/min)

Accuracy

±(ppm of reading + ppm of range) [1]

Distortion

( % )

Class 0.05

Class 0.02

10 kV

100 mV

0.01

300 + 200

120 + 80

≤ 0.5

35 kV

100 mV

0.01

300 + 200

120 + 80

≤ 0.5

Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%).

• Output range: 1 kV~35 kV,6 digits true RMS,max burden: 600 VA

• Frequency range: 45.000 Hz~65.000 Hz,uncertainty:0.01Hz

• Short-circuit and overload protection

• Harmonic function(option): 2nd~11th harmonic, 0~25% adjustable


6. General Specifications

Power Supply

AC (220 ± 22) V,(50 ± 2) Hz

Temperature

Working temperature:0℃~45℃;

Storage temperature:-20℃~70&℃

Humidity

Working humidity:< 80% @ 30℃,< 70% @ 40℃,< 40% @ 50&℃

Storage humidity:(20%~80%) R·H,non-condensing

Interface

RS232 interface


7. Ordering Information

TUNKIA TD2410 High Voltage Meters Calibration Device

Previous: 
Next: 

SIGN UP FOR NEWSLETTER

Get latest deals, newsletter and technical information

INFORMATION

CONTACT US

 WhatsApp:+86-18773180215
     Wechat: 18773180215
 Phone: +86-18773180215
 E-mail: global@tunkia.com
E-mail: Eugene@tunkia.com
 Address: NO.16 Panpan Road, Changsha
Copyright  2021 TUNKIA Co., Ltd. All Rights Reserved. Supported by leadong.comSitemap