loading

TS2000 Rotational Magnetic Properties Measuring System for Electrical Steel

TS2000 is a special instrument for measuring orientation (full Angle) magnetic properties of electrical steel sheet. The magnetic properties of the specimens can be measured within a certain frequency range from 0° to 360°.

  • TS2000

  • TUNKIA

Availability:

TS2000 Rotational Magnetic Properties Measuring System for Electrical Steel


TS2000 Rotational Magnetic Properties Measuring System for Electrical Steel

*Above picture is only for reference, subject to actual delivery


1. Summary

TS2000 is a special instrument for measuring orientation (full Angle) magnetic properties of electrical steel sheet. The magnetic properties of the specimens can be measured within a certain frequency range from 0° to 360°.


2. Features

  • Electrical parameters calibration function.

  • Operators only need to insert a single specimen into the permeameter to test.

  • Measurement of magnetic field by M.C and H-Coil

  • Slowly reduce amplitude demagnetization for specimen.

  • Software controlled automatic measurement.

  • The software has been configured with multiple measurement schemes

  • Automatic calculates magnetic parameters and curves.

  • Modular design, easy to upgrade or maintenance.


3. Applications

TS2000 Rotational Magnetic Properties Measuring System for Electrical Steel Applications



4. Specifications

4.1 Specification of Electrical Parameter

Output Capacity


Excitation Voltage

0.1 Vrms ~ 200 Vrms

Excitation Current

0.5 mApk ~ 15 Apk

Frequency Range

45 Hz ... 400 Hz

Effective Excitation Power

0~500 VA

Measuring Indicators

Frequency Range (Hz)

Voltage And Current Effective Value

Optimal Measurement Uncertainty (k=2)

Active Power

Optimal Measurement Uncertainty (k=2)

45 ≤ F ≤ 65

0.05%*RG ⑤

0.1%*FS ⑥

65 < F ≤ 400

0.05%*RG ⑤

0.15%*FS ⑥

• Note ⑤: RG is the range value; ⑥: FS = voltage range value × current range value, when the phase angle is within 75°.


4.2 Specification of Magnetic Parameter

Typical Frequency

Magnetic Parameter

Typical Testing Points①

Optimal Measurement Uncertainty (k=2)

Best Repeatability

50Hz / 60Hz

Hm

10 A/m...5 kA/m

0.5%

0.2%

KJ④

J10 ②... J5000

0.5%

0.2%

KPs⑤

P0.1 ③... P1.7 ( P1.5 )④

1.0%

0.5%

400 Hz

Hm

10 A/m...1 kA/m

1.0%

0.3%

KJ④

J10 ②... J1000

1.0%

0.3%

KPs⑤

P0.1 ③... P1.7 ( P1.5 )④

2.0%

1.0%

• Test conditions: (23 ± 5)℃, the specimen should be demagnetized before testing.

• Accuracy of measurement value: 1% to 2% deviation will occur depending on the way and position of loading the specimen in the measurement frame.

• Repeatability of measurement value: The same specimen is placed in the specified position and kept without moving, and measured repeatedly 10 times. Calculation method: Repeatability = (maximum value - minimum value) ÷ average value × 100%.

• Note ①: The table only gives reference test points. The Hm, Jm / Bm, Ps and other indicators that the sample can achieve during actual testing are related to the performance, quality, test frequency, test current, load voltage, etc. of the material.

• Note ②: J10 refers to the magnetic polarization strength when the magnetic field H is 10 A/m, and so on. The brackets are typical test points for non-oriented steel.

• Note ③: P0.1 refers to the total loss value when the magnetic flux B is 0.1 T. The maximum test point is the maximum test point marked at 0°.

• Note ④: KJ = J_δ / J_0°, the test specimen is a circular specimen.

• Note ⑤: KPs = Ps_δ / Ps_0°, the test specimen is a circular specimen.


4.3 TS2000-SST Full Angle Single Sheer Tester

Magnetic Field Strength

10 A/m ~ 5 kA/m

Operating Frequency

45 Hz ~ 400 Hz

Magnetic Field Measurement

Current measurement method (M.C) and magnetic field coil method (H-Coil method)

Magnetic Direction

0 ~ 360°

Operating Temperature

0℃ ~ 40℃


4.4 Specimen

Specimen size

300 mm (W ) × 300 mm (L)

Specimen thickness

0.18mm ~ 0.65mm

Specimen shape

Square or round


4.5 General Specification

• Power supply: AC (220 ± 22) V, (50 ± 2) Hz

• Working environment: 0 ℃ ~ 40 ℃, 20% ~ 85% R·H, no condensation

• Air supply requirements: Compressed air pressure not less than 0.6 MPa, air consumption not less than 100 L/min

• Air source can be purchased from an air compressor or installed at the site


5. Configuration List

S/N

Name

Qty

Config.

Remark

1

TS2000 Excitation and Measurement Host

1

Standard


2

TS2000-SST Full Angle Single Sheet Tester

1

Standard

≤ 400Hz,10 A/m~5 kA/m

3

Testing Software

1

Standard


4

Instructions and Certificate of Conformity

1

Standard


5

Test Leads, Supplied Accessories

1

Standard


8

Work Bench

1

Optional

Third party product

9

Sartorius Electronic Balance

1

Optional

0.1 ~ 3.2 kg, accuracy: 0.1 g

10

Computer

1

Optional

Third party product

11

Printer

1

Optional

Third party product

• Note: The above is for reference only. The specific configuration list shall be subject to the technical agreement.

Previous: 
Next: 

SIGN UP FOR NEWSLETTER

Get latest deals, newsletter and technical information

INFORMATION

CONTACT US

 WhatsApp:+86-18773180215
     Wechat: 18773180215
 Phone: +86-18773180215
 E-mail: global@tunkia.com
E-mail: Eugene@tunkia.com
 Address: NO.16 Panpan Road, Changsha
Copyright  2021 TUNKIA Co., Ltd. All Rights Reserved. Supported by leadong.comSitemap